What are the daily maintenance of high and low temperature Thermal Shock Chamber?
SONACME 2021-10-14 10:30:34
1. The high and low temperature Thermal Shock Chamber should be fixed to clean the condenser every 3 months. For the refrigeration system adopts air-cooled cooling, the condensing fan should be regularly inspected and the condenser should be decontaminated and dust-removed to ensure good ventilation. Thermal performance.
2. High and low temperature Thermal Shock Chamber If the low temperature is used for a long time, after one cycle, the temperature should be set to 110 degrees, and the door should be opened for two hours for defrosting treatment. At the same time, it should be insisted that after each test, set the temperature near the ambient temperature, work for about 30 minutes, then cut off the power supply, and wipe clean the inner wall of the working chamber.
3. High and low temperature Thermal Shock Chamber should clean the evaporator regularly. Because the cleanliness of the test products are different, under the action of forced air circulation, a lot of small particles such as dust will condense on the evaporator, and it should be cleaned regularly.
4. High and low temperature Thermal Shock Chamber waterway and humidifier cleaning, if the waterway is not smooth, humidifier fouling will easily cause the humidifier to dry out, which may damage the humidifier, so the waterway and humidifier must be cleaned regularly.